- 177 -
Give balance serial number
Format:
NB CR LF
Accessible responses:
NB_A_”x” CR LF
NB_I CR LF
- command understood, returns serial number
- command understood but not accessible at this
moment
x – serial numer of the device (inserted in between inverted commas)
Example
Command:
NB CR LF
– give serial number
Response:
NB_A_”1234567”
– serial number of the device – ”1234567”
Value release
Format:
SS CR LF
Accessible responses:
SS_OK CR LF
- command understood, execution in progress
Command’s function is similar to function of button PRINT located
on an overlay,
Internal calibration
Format:
IC CR LF
Accessible responses:
IC_A CR LF
IC_D CR LF
- command understood, execution in progress
- calibration completed
IC_A CR LF
IC_E CR LF
- command understood, execution in progress
- range exceeded, time limit for awaiting for a stable result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal calibration of the balance
Format:
IC1 CR LF
Accessible responses:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified balance
IC1_OK CR LF - command carried out
Operation is disabled for verified balances.
For non-verified balances the command inhibits internal calibration until it is
enabled via IC0 command or until the balance is turned off. The command does
not modify settings specifying calibration start.