1260-00C User Manual
Local Command Set 4-45
Command
Syntax:
A16 <A16 address>, <word value>
Action: Write
<word value>
to
<A16 address>
.
Example:
Write A502h to VXI A16 address 4305h.
A16 #h4305, #hA502
A16?
Purpose:
Read word value from VXI A16 address space.
Query
Syntax:
A16? <A16 address>
Response:
Program response:
<word value>
<CRLF>
Console response:
Value <hex word value> (<word value>)
read from A16 address <A16 hex
address>
(<A16 address>)<CRLF>
where
<word value>
is in decimal integer format,
<hex word value>
is in C language
hexadecimal format,
<A16 hex address>
is in C
language hexadecimal format, and
<A16
address>
is in decimal integer format.
Example:
Read the ID register of Logical Address 16.
A16? #hc400
A24
Purpose:
Write a 16-bit value into VXI A24 space.
Command
Syntax:
A24 <A24 address>, <word value>
Note that the
<A24 address>
has a valid range
of 2097152 to 14680062
(#h200000 to
#hE7FFFE).
Action:
Write
<word value>
to
<A24 address>.
Example:
Write the value A502h to VXI A24 address
504305h.