1260-00C User Manual
Operation 3-22
Indications for Message-
Based Device Operation
FAILED, TEST, and ON LINE LEDs. The LED indications are
shown in Table 3-13. A successful system start-up sequences
through the first five states. The point of failure is indicated for
states in which the FAILED LED is lit for an extended period of
time. The LED indications are identical for Non-Slot 0
Message-Based device and Slot 0 Message-Based device
operation.
Table 3-13, Front Panel LED Indications For Message-Based Device Operation
Sequence
FAILED
TEST
ON LINE
State
Point of
Failure
1
OFF
OFF
OFF
No power
2
ON
OFF
OFF
In self-initialization
Failed before self-
test
3
ON
ON
OFF
In self-test
Failed in self-test
4
OFF
ON
ON
Waiting for BNO
5
OFF
OFF
ON
Online
ON
OFF
ON
Failed
Failed while online
OFF
ON
OFF
In non-volatile
configuration, or
diagnostics mode
Slot 0 Message-
Based Device
Configuration
To configure the 1260-00C for Slot 0 Message-Based operation,
use the following procedure:
a. Enable the VXIbus Slot 0 functions.
b. Set the model code of the 1260-00C to be configured for
Slot 0 operation using the non-volatile configuration mode.
c. Set the logical address to a non-zero value with an
appropriate Servant area size. If the logical address is set
to FFh, the 1260-00C will participate in dynamic
configuration. Otherwise, the 1260-00C is a static
configuration device.
d. Set the various switches, jumpers and non-volatile
configurations as shown in Table 3-14.
Table 3-14, Slot 0 Message-Based Device Operation Switch and Jumper Settings