1260-00C User Manual
Local Command Set 4-7
Syntax:
CONF
Example:
Reboot into non-volatile configuration mode.
CONF
ConsoleEna
Purpose:
Enable or disable the RS-232 port as the console.
When the RS-232 port is disabled as the console, a CI can take
control of the serial port.
Command
Syntax:
ConsoleEna <Boolean>
Action:
If <
Boolean
> is TRUE,
ConsoleEna
sets the
RS-232 port to be a local command set input.
If
<Boolean>
is FALSE,
ConsoleEna
disables the
RS-232 port connection to the local command set.
Note that once the console has been disabled, it
must be re-enabled from a command source other
than the RS-232 port (such as the GPIB port).
Examples:
Disable console.
ConsoleEna 0
Enable console.
ConsoleEna 1
ConsMode
Purpose:
Enable or disable the console data mode.
Command
Syntax:
ConsMode
<Boolean>
Action:
If
<Boolean>
is TRUE,
ConsMode
enables console
format responses for the command source issuing
the command.
If
<Boolean>
is FALSE,
ConsMode
disables
console format responses for the command source
issuing the command.
The console response mode applies only to the
response path connected to the
ConsMode
command source. For example, disabling the
console response mode from the GPIB port does