1260-00C User Manual
Local Command Set 4-10
Response:
Program response:
<memsize>
<CRLF>
where
<memsize>
is the amount of installed RAM,
in kilobytes.
Console response:
This 1260-00C has <expression> of RAM
installed onboard.
<CRLF>
where
<expression>
is the amount of installed
RAM.
ProgMode
Purpose:
Enable or disable the program data mode.
Command
Syntax:
ProgMode
<Boolean>
Action:
If
<Boolean>
is TRUE,
ProgMode
enables
program format responses for the command source
issuing the command.
If
<Boolean>
is FALSE,
ProgMode
disables
program format responses for the command source
issuing the command.
The program response mode applies only to the
response path connected to the
ProgMode
command source. For example, disabling the
program response mode from the GPIB port does
not affect the response mode on the serial port.
Examples:
Disable program format responses.
ProgMode 0
Enable program format responses.
ProgMode 1
WordSerEna
Purpose:
Assign control of the 1260-00C physical Word
Serial registers to an onboard logical address
(1260-00C command interpreter or code
instrument).