1260-00C User Manual
Local Command Set 4-56
Bit General Configuration Modes
1
1 = Pass trigger through asynchronously
0 = Synchronize with next CLK10 edge
2
1=Participate in SEMI-SYNC with
external trigger acknowledge protocol
0 = Do not participate
All other values are reserved for future expansion.
Response:
Program response: 0
Console response:
Trigger assertion configuration (line
= <line text>, mode = <mode>)
complete.
<CRLF>
where the meaning of
<line text>
corresponds
to the value of
<line>
:
Value of
<line>
Value of
<line text>
0 to 7
TTL <line>
8 to 9
ECL (<line> - 8)
ffffh
GENERAL CONFIG
Example 1: Configure all TTL/ECL trigger lines generally to
synchronize to the falling edge of CLK10 (as
opposed to the rising edge).
TrigAsstConf -1, 1
Example 2: Configure TTL trigger line 4 to synchronize to
CLK10 for any assertion method and do not
participate in SEMI-SYNC.
TrigAsstConf 4, 0
TrigCntrConf
Purpose:
Configure the TIC chip internal 16-bit counter.
Query
Syntax:
TrigCntrConf <mode>, <source>, <count>
where
<mode>
specifies the configuration mode:
Value Mode
0
Initialize the counter
2
Reload the counter leaving enabled
3
Disable/abort any count in progress
<source>
specifies the trigger line to configure as