1260-00C User Manual
Local Command Set 4-59
used as feedback and left tri-stated for use as a
mapped input via
MapTrigTrig
.
TrigExtConf 40, 16
Example 3 : Configure external line 48 (GPIO 8) to be used as
feedback for use as a cross-point switch input and
output via
MapTrigTrig.
TrigExtConf 48, 1
TrigTickConf
Purpose:
Configure the TIC chip internal dual 5-bit tick timers.
Query
Syntax:
TrigTickConf <mode>, <source>,
<tcount1>, <tcount2>
where
<mode>
specifies the configuration mode:
Value Mode
0
Initialize the tick timers (rollover mode)
1
Initialize the tick timers (non-rollover mode)
2
Reload the tick timers leaving enabled
3
Disable/abort any count in progress
and the value of
<source>
is the trigger line to
configure as input to counter:
Value Trigger Line
40 to 49
External source/destination (GPIO
0 to 9)
70
CLK10
71
EXTCLK
and the values of
<tcount1>
and
<tcount2>
are
the number of input pulses (as a power of two) to
count before asserting TICK1 output or TICK2
output (and terminating the tick timer if configured
for non-rollover mode).
Response:
Program response: 0
Console response:
TICKs configured (mode = <mode text>, source
= <source text>, t1 = tcount1, t2 =
tcount2).
<CRLF>
where the meaning of
<mode text>
corresponds
to the value of
<mode>:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com