1260-00C User Manual
Local Command Set 4-54
ECL, or external trigger line.
Query
Syntax:
SrcTrig <line>, <protocol>, <timeout>
where the value of
<line>
corresponds to the
trigger line to source:
Value Trigger Line
0 to 7
TTL trigger lines 0 to 7
8 to 9
ECL trigger lines 0 to 1
40 to 49
External source/destination
(GPIO 0 to 9) (supports ON,
OFF, START, STOP, and
SYNC protocols only)
50
TIC counter
(supports SYNC and
SEMI-SYNC protocols only)
60
TIC TICK timers
(supports SYNC and
SEMI-SYNC protocols only)
the value of
<protocol>
specifies the protocol to
use:
Value Protocol
0
ON
1
OFF
2
START
3
STOP
4
SYNC
5
SEMI-SYNC
6
ASYNC
7
SEMI-SYNC and wait for
Acknowledge
8
ASTNC and wait for Acknowledge
ffffh
Abort previous Acknowledge
pending (5 and 6)
and
<timeout>
is the timeout value in milliseconds
Response:
Program response: 0
Console response:
Trigger sourcing (line = <line text>,
protocol = <protocol text>)
complete.
<CRLF>
where the meaning of
<line text>
corresponds
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com