Operation 3-21
appropriate Servant area size using the non-volatile
configuration mode. If the logical address is set to FFh, the
1260-00C will participate in dynamic configuration.
Otherwise, the 1260-00C is a static configuration device.
d. Set the various switches, jumpers and non-volatile
configurations as shown in Table 3-12.
Table 3-12, Non-Slot 0 Message-Based Device Operation Switch and Jumper Settings
Jumper/Switch
Position
Function
Switch S15
OFF
If S15 is ON, the 1260-00C
sources CLK10 at the front
panel EXT CLK connector.
Switch S22
OFF
VXI BTO disabled.
Switch S23
OFF
Bus arbiter and SYSCLK
disabled. CLK10 receiving
form backplane.
Switch S24
OFF
MODID pulled down.
Logical Address
See Chapter 5.
Logical address is not equal
to 0. Set in non-volatile
configuration or by using DIP
switch SW1.
Non-Slot 0 Model Code
See Chapter 5.
Model code is set to the Non-
Slot 0 value. Set in non-
volatile configuration.
Servant Area Size
See Chapter 5.
Set appropriate Servant area
size. Set in non-volatile
configuration.
Non-Slot 0 Message-
Based Device Operation
At start-up, a 1260-00C configured as a Non-Slot 0
Message-Based device performs its self-tests. It then waits until it
receives its Device Grant and Begin Normal Operation Word Serial
commands. The RM grants any logical addresses to the
1260-00C that reside within its Servant area. When it responds to
the Begin Normal Operation command, the 1260-00C enters its
normal mode of operation
.
Front Panel LED
The 1260-00C indicates the progress of its self-test with the