1260-00C User Manual
Local Command Set 4-51
where the value of
<line>
corresponds to the
trigger line or counter/tick source:
Value Trigger Line
0 to 7
TTL trigger lines 0 to 7
8 to 9
ECL trigger lines 0 to 1
50
TIC counter
60
TIC TICK1 tick timer
Response:
Program response: 0
Console response:
Trigger handler (line = <line text>):
DefaultTrigHandler()
.<CRLF>
where the meaning of
<line text>
corresponds
to the value of
<line>
as follows:
Value of
<line>
Value of
<line text>
0 to 7
TTL <line>
8 to 9
ECL (<line> - 8)
50
TCNTR
60
TICK 1
Example:
Get the address of the trigger interrupt handler for
TTL trigger line 4.
GetTrigHndlr 4
MapTrigTrig
Purpose:
Map a specified TTL, ECL, Star X, Star Y, external
connection (GPIO), or miscellaneous signal line to
another.
Query
Syntax:
MapTrigTrig <srcTrig>, <destTrig>,
<mode>
where
<srcTrig>
is the source line to map to a
destination, the value of
<destTrig>
corresponds
to the destination line to map from the source:
Value Trigger Line
0 to 7
TTL trigger lines 0 to 7
8 to 9
ECL trigger lines 0 to 1
40 to 49
External source/destination (GPIO 0 to 9)
40
Front panel In (connector 1)
41
Front panel Out (connector 2)
42
Front panel In (unbuffered)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com