8.6.4
DULCOMETER
®
D1C Measured Variable pH and DULCOTEST
®
PHEP or
PHER Sensor for pH, Maintenance
Controller
The
DULCOMETER
®
controller type D1C for the measured variable pH is maintenance-free.
Sensor
IMPORTANT
• The sensor is to be regularly serviced to avoid any excess metering caused by a sensor
failure!
• Observe the valid national regulations for maintenance intervals!
• Do not touch the electrodes or bring into contact with greasy substances!
Maintenance interval
monthly
Maintenance
Work
The sensor should regularly (once a month) undergo a visual check and be cleaned if necessary.
If deposits on the glass electrode withstand cleaning with a soft, moistened cloth, the following
cleaning agents may be used.
Kind of deposit
Agent and duration of application
General deposits
Non-abrasive household cleaner
Scale or metal hydroxides
Diluted hydrochloric acid (approx. 0.1-3 %, 1-5 minutes)
Oil, grease
Solvents, like alcohol or acetone
Biofouling
Mixture of diluted hydrochloric acid and pepsin, several hours
Solvents (e.g. acetone) must not be used to clean electrodes as
they can demage the plastic stems.
It is essential that the sensor are rinsed thoroughly after having been cleaned.