Requirements:
n
1 weighing instrument which can be read sufficiently accurately
n
1 vessel with feed chemical
n
the suction line is fed, bubble-free into the measuring vessel.
1.
Zero the weighing instrument.
2.
Change from the menu item
„calibration stroke feedback“
using the
[arrow keys]
to the menu
item
„calibration quantity“
and press key
[P]
.
ð
The menu item
„quantity corr. factor“
appears.
3.
Change from the menu item
„quantity corr. factor“
using the
[arrow keys]
to the menu item
„quantity meas. Value“
and press key
[P]
.
ð
The menu item
„actual mass“
appears.
4.
Start the metering pump.
Select the number of strokes so that the reading error at the weighing instrument (half of
the smallest readable mass graduation divided by the metered total mass) is smaller
than the Dulcoflow measurement error.
5.
Stop the metering pump.
6.
Read off the weight from the weighing instrument.
7.
Press key
[P]
.
ð
The menu item
„nominal mass“
appears.
8.
Adjust the value in the menu item
„nominal mass“
using the
[arrow keys]
to the read-off
weight and press the key
[P]
.
ð
The menu item
„corr. factor“
appears. It shows the calculated correction factor.
9.
Press key
[P]
, to return to the continuous display.
It is also possible to manually calculate the correction factor and enter it directly under
„entry“
-
„correction factor“
.
8.4.4 "Zero set" main menu
main menu
zero set
In the main menu "Zero set", the total quantity
and the strokes, which are displayed in the con‐
tinuous display, can be set to "zero" independ‐
ently of each other.
Settings
40