About setpoint and tolerances
B0232
P
P
P
P
P
P
P
calibration
stroke feedback
mass (stroke)
0.347 g
stroke feedback
set-point
mass (stroke)
0.347 g
lower tolerance
-010.00 %
tolerance
tolerance (%)
upper tolerance
+020.00 %
Specify the allowed range for the mass per stroke m/H via the setpoint of the lifting volume m/H and
tolerances in %:
1.
Follow the menu path
„calibration
è
stroke feedback
è
set-point“
and press the key
[P]
.
ð
The menu item
„volume (stroke)“
displays the currently stored setpoint.
2.
Start the metering pump.
ð
The actual measured value is displayed.
3.
Turn the stroke length adjustment knob until the desired setpoint is displayed.
4.
Press key
[P]
.
ð
The displayed measured value is saved as a setpoint and the menu item
„tolerance
tolerances“
appears.
The setpoint is valid as 100%.
5.
Press key
[P]
.
6.
Set the
„lower tolerance“
with the
[arrow keys]
and press the
[P]
.
7.
Set the
„upper tolerance“
with the
[arrow keys]
and press the
[P]
.
ð
The continuous display appears again.
8.
Stop the metering pump.
The desired setpoint can also be entered, without having to use the stroke adjustment dial or
the pump having to be running, directly under „mass (stroke)“ using the [arrow keys] .
Settings
33