8.4.2.3 "Current output" menu
(identity code characteristic "signal output" "1")
P
P
P
P
P
P
P
P
B0234
set-up
current output
current range
4...20 mA
current output
signal current
flow unit
mL/h
0/4 mA value
0.68 mL/h
20 mA value
897.65 mL/h
current output
error current
error current
bubbles
In the
„current output“
menu the standard signal output (mA) can be set either to output the instanta‐
neous flow or as a feedback message.
In the
„signal current“
menu a setting can be
made to determine how the DulcoFlow
®
outputs
the instantaneous flow via the current output to
an external device.
In the
„error output“
menu a setting can be
made to determine whether the DulcoFlow
®
outputs an error signal via the current output to
an external device (23 / 3.6 mA).
Settings
28