Page
90
October 2014
For example
: measure a 16 bit data bus signal, bus will appear data of
"0X99" and continuous time is 50ns. We set BUS0 to include 16
measurement channels and sample rate as "100M", trigger setting as
duration trigger. Setting steps as below:
Press "
Trigger MENU
" and menu appears.
Press "
F1
" till trigger mode display as "
Data Width
".
Press "
F2
" till signal resource display as "
BUS0
"
Press "
F3
" and data in code type as red background and green digital
indicator on, then insert data value "
0X99
". Press "
F3
" again and red
background disappeared. Repeat operation of step D if setting error or
need to be modified.
Press "
F4
" and duration in menu setting window correspond to
F4
display red background. Turn "
CH1 Volts/Div
" knob and set duration
as "
50ns
".
Press "
F5
" till qualifier display as "
>=
"
Turn "
Trigger level
" adjust knob or press "
set 50%
"till trigger
position display as "
NEXT T POS = 50%
".
Then duration trigger setting finished (refer to Fig. 86) and ready for data
acquisition.
Figure 86.
Data width trigger.