October 2014
Page
89
►
Data width queue trigger:
Make BUS as trigger source and continuous duration in BUS as trigger
condition to generate trigger. Duration trigger menu refer to Fig. 85.
Figure 85.
Duration trigger menu.
Duration trigger function as below:
Function
Setting
Instruction
Source
BUS0
∼
BUS3
Select the trigger source from BUS0~BUS3.
Code
0x0000
∼
0xffff
“HEX”
0
∼
65535 “DEC”
Can be set discretionarily between 0x0000 and
0xffff “HEX” or between 0 and 65535 (DEC)
according to the bus and code setting.
Duration
10ns
∼
50s
The duration width can be set from 10ns
(1-2-5) to 10.00us (1-2-5) according to the
sampling rate from high to low.
Qualifier
>=
<=
Trigger occur on condition that the bus value is
equal to the code type and the duration is more
than or equal to the duration width.
Trigger occur on condition that the bus value is
equal to the code type and the duration is less
than or equal to the duration width.