3.6
Logic Analyzer
3.6.1
How to set sampling system
Sampling system is to set sample rate, storage depth and filter. Different
sampling setting will result in different measure results. In the same storage
depth, the higher sample rate set, the shorter the continuance time for signal will
be. If the sample rate set too low, narrower pulse signal may get lost. So the
sample rate and storage depth should be set correctly according to the actual
measuring signal. Press "
E
(
ACQU
)"and display menu as the Fig.73:
Figure 73.
Sampling menu.
Sample menu function list:
Function
Setting
Instruction
Sample rate
20S/s to 1GS/s
Storage
depth
Low memory
General
Deep memory
Storage depth of 16K
Storage depth of 256K
Storage depth of 4M
Digital Filter
None
1
2
Filter closed
Filter one pulse of sample width
Filter two pulse of sample width