Here are the procedures for carrying a self-calibration.
Disconnect all inputs, including probes and wires etc.
Press
UTILITY
to activate Function menu.
Press
F3
next to Do Self-Cal.
Now there is a yellow window popping up to ask for a confirmation. Press F3
again, the Oscilloscope starts the self-calibration and yellow window remains the
calibration is done.
Please be ware that no contacts to any inputs channels
till the calibration is done.
►
SYS STAT (System State)
Press the
F1
menu selection button and choose "
SYS STAT
" item. The menu
pops up in the screen as Fig. 61.
Figure 61.
SYS STAT Menu.
Function Menu
Setting Description
Horizontal
Show the horizontal parameter of the channel.
Vertical
Show the vertical parameter of the channel.
Trigger
Show the parameters of the trigger system.
Misc
Show the serial number and edition number.
After entering into the SYS STAT menu, choose the corresponding function, with
the corresponding parameters shown in the screen. If press the
F1
menu
selection button and choose the function item "
Horizontal
", the Horizontal
System State will be displayed in the screen. Press any other function button and
exit from the SYS STAT menu (see Fig. 62).