XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY
display format is selected, Channel 1 is displayed in the horizontal axis and
Channel 2 in the vertical axis; the oscilloscope is set in the un-triggered
sample mode: the data are displayed as bright spots and the sampling rate
is 1MS/s and can not be changed.
The operations of all control knobs are as follows
:
The Vertical
VOLTS/DIV
and the
VERTICAL POSITION
knobs of
Channel 1 are used to set the horizontal scale and position.
The Vertical
VOLTS/DIV
and the
VERTICAL POSITION
knobs
of
Channel 2 are used to set the vertical scale and position continuously.
NOTE: The following functions can not work in the XY Format
:
Reference or Any calculated 1 wave form.
Cursor Function.
All Time base controls.
All Trigger controls.
►
Use XY Format
Apply input signals to both CH1 and CH2.
Press the
DISPLAY
to activate
DISP Set
menu.
Press
F1
to set Vectors as type,
F2
to set Infinite for persist and
F3
to
set XY as Format as shown as Fig. 55
Figure 55.
Display in XY Format.