![Promax OL-612 Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/promax/ol-612/ol-612_user-manual_1616414121.webp)
Page
116
October 2014
►
No Display Responses to the Push-down of RUN/STOP.
Check whether Normal or Signal is chosen for Polarity in the TRIG MODE
menu and the trigger level exceeds the wave form range.
If it is, make the trigger level is centered in the screen or set the trigger
mode as Auto. In addition, with the
AUTOSET
button pressed, the setting
above can be completed automatically.
►
After the AVERAGE value sampling is set in the ACQU MODE or the
longer duration is set in the DISP MODE, the display rate is slowed
down.
It is a normal phenomenon.