Risk of injury: Never hold the sample by hand during operation. It has to be
fixed with an appropriate means, e.g. double sided adhesive tape.
Scratch tests are carried out in machine mode „Applicator“ in moving mode
„Single“. Go to the menu on the touchscreen -> select Setup, set the
machine mode to „Applicator“, visible through an arrow
on the
touchscreen.
•
Place the device on a solid, even surface.
•
Level the apparatus using the included spirit level. To do so, place the spirit
level on the sample tray and adjust the levelling feet (8) until the Proceq ZAA
2600 is levelled perfectly. The glass plate must be firmly in place.
•
Place the sample.
-
For very thin samples use the glass plate with the glass side facing up.
-
For thicker samples use the glass plate with the printing blanket facing up.
In order to get straight scratch marks fix the samples with a double sided
adhesive tape.
•
Move the tool holder (18) either by tapping the symbol
or by hand to the
back part of the sample.
•
Pre-assemble the chosen scratching tool (see chapter 7.2
scratching and mar resistance sets