Page 47
12.4 Replacing the fuse
If the apparatus cannot be switched on this could be due to a defect fuse. You can
replace the fuse yourself by opening the fuse holder (13) on the back of the
apparatus between the main switch (14) and the power supply (12) carefully using a
screw driver or a similar tool.
Before replacing the fuse of the Proceq ZAA 2600, always switch off the
apparatus and unplug it.
Replace the microfuse (14) only by an equivalent fuse of the same type
with the same specification.
Observe that the inscription at the connection
caps of the fuse corresponds to one of the following designations:
•
T 0.8 A / 250 V
•
T 800 mA / 250 V
•
T 800 / 250 V.
•
Clamp the new fuse into the fuse holder (13) and put it back in its place.
Should it still be impossible to switch on the apparatus, please contact
Proceq SA
or your authorized Proceq agent.
12.5 Inspection
The instrument should be checked for proper condition by a qualified electrician at
an interval of 2 years. Alternatively, this test may also be performed by Proceq SA.