Page 39
8.6 Reset
With the „Reset“ button the cycle
counter (double stroke counter) will be
reset. This cannot be undone.
If the „Reset“ button has been tapped
this message appears which has to be
either confirmed or rejected.
8.7 Status messages
8.7.1 Overload
"Motor has stopped because of
overload. “This message is only an
information, you can tap either on “OK”
or move forward with
. Should
the message be displayed repeatedly
during a procedure, the resistance is
probably too high. A possible solution
could be to increase the acceleration
mode. Select “Setup” and “Motor
Settings”, then “Acceleration mode”
and tap „Elevated“ in the menu.
8.7.2 End switch
“Motor has stopped because end
switch was reached“. The preset stroke
length exceeds the position of the end
switch. It is recommended to always
move the end stop (2) completely to
the front and the start stop (1)
completely to the back part of the
device.
Solution: Move the end stop (2)
completely to the front of the device
and press „OK“ or „Scrubbing“
afterwards.