14 Further information
■
67
14.2.4 Automatic calibration
For an automatic calibration with the built-in calibration weight “SET
CALIBRATION MODE AUTO” must be selected in the configuration
menu (see Chapter 5.6 “Calibration functions”).
The balance now calibrates itself automatically every 24 hours and/or
after each temperature change of 3 degrees Celsius, depending on the
definition in the configuration menu “SET CALIBRATION AUTOCAL”.
The time of the automatic calibration is as determined in the configu-
ration menu under “SET CALIBRATION AUTOCAL. -TIME n h”.
(e.g. 6 h for 06.00 o'clock in the morning).
NOTE
For the automatic calibration by time and by time/temp. the date
and time of the balance must first be correctly set (see Chapter 5.9
“Date and time”).
The calibration can also be effected manually any time while auto-
calibration is activated.
The automatic calibration then takes place only if no weight is pla-
ced on the pan for at least five minutes.
It is recommended, that the time for the auto-calibration be set at
a time outside the normal business hours (for example, in the early
morning).