14 Further information
66
Then proceed as follows:
14.2.3 Internal calibration
For an internal calibration with the built-in calibration weight “SET
CALIBRATION MODE INTERNAL” must be selected in the configura-
tion menu (see Chapter 5.6 “Calibration functions”).
Then proceed as follows:
• Switch to “BALANCING” with the «
MODE
»-key
• Press «
T
» until “CALIBRATION” is shown.
• The calibration is finished after a certain period of time.
• Switch to “BALANCING” with the
«
MODE
»-key
cALIBRation
• Press «
T
» until “CALIBRATION”
appears.
- - 0000
9
• The balance carries out a Zero
measurement (0000 g is shown
flashing)
- - 372
9
• After the zero measurement the
display flashes with the previous-
ly entered calibration weight
• Place the calibration weight on
the pan
- - 372
9
• The display flashes rapidly
• Calibration is complete when the
display stops flashing (the exact
weight is shown)
+ 372.15
9