Precisa XB series Скачать руководство пользователя страница 15

2 Your balance

   

13

 The balance shows the following labels:

Fig. 2.3

Balance labels

2.3

Data and parameters

The Precisa balances of Series XB are divided into five main-groups A, 

C, D, G and M. The letter in the name corresponds to the design 

specification (e.g. A = analytical balance, M = MilIigramme balance) 

the number before it corresponding in each case to the maximum 

allowable load (in grammes).
The allowable weighing range, the calibration value and the readability 

of the balance are printed on the casing (see Tab. 2.1 List of appliance 

labels) and are therefore not presented here. 

6

7

XB 2200 C

Max. 2200 g

Min. 0,5 g

e=0,1g

d=0,01g

2HA?EI=

1

2

3

4

5

PRINT

MODE

ON/OFF

swissmade

REF

T

Содержание XB series

Страница 1: ...Operating Instructions 350 8105 000c1 Precisa Balances Series XB ...

Страница 2: ......

Страница 3: ... ch http www precisa com Refer to our website for information about local customer service cen ters and details of their addresses Copyright These operating instructions are protected by copyright All rights reserved No part of the operating instructions may be reproduced processed duplicated or published in any form by photocopying microfilming reprinting or other process in particular electronic...

Страница 4: ... accentuated by printing in semi bold script ON OFF In the explanation of the operating steps the appropriate display for the current operating step is shown for clarity at the left alongside the list of operating steps For the labelling of dangers and advice please see Chapter 1 Safe ty Guarantee card A guarantee card which was filled in by your Precisa dealer before handing over the balance is e...

Страница 5: ...15 3 1 Unpacking the balance 15 3 2 Scope of delivery 16 3 3 Assembly of the balance 17 3 4 Choice of a suitable location 17 3 5 Checking the mains voltage 18 3 6 Levelling the balance 18 3 7 Calibration of the balance 19 3 8 Dual Range and Floating Range balances 19 3 9 Standardized balance 20 3 10 Under floor weighing 20 4 Modes of operation and operating 22 4 1 Switching on the balance 22 4 2 A...

Страница 6: ...ast 30 5 4 Selecting the weight unit 31 5 5 Print functions 32 5 6 Calibration functions 33 5 7 Weighing mode 34 5 8 Interface functions 35 5 9 Date and time 36 5 10 Password protection 36 5 11 Anti theft encoding 37 6 Working with the application menu 38 6 1 Structure of the application menu 38 6 2 Selecting a working program 39 6 3 Changing application weighing mode 39 6 4 Application UNITS 40 6...

Страница 7: ...on of password protection 56 10 2 Selection of an application program 57 10 2 1 Setting for Counting by weighing 57 11 Maintenance and servicing 59 12 Transport storage and disposal 60 12 1 Transportation and shipping of the balance 60 12 2 Storage of the balance 60 13 Accessories 61 13 1 Accessories 61 14 Further information 63 14 1 Notes on the Weighing mode 63 14 1 1 Set Weighing mode FLOATINGD...

Страница 8: ... Contents 6 14 2 3 Internal calibration 66 14 2 4 Automatic calibration 67 15 Configuration menu tree 68 16 Application menu tree 71 17 Configuration menu tree 71 ...

Страница 9: ...e power adaptor make sure that the operating voltage stated on the power adaptor agrees with the mains voltage If not please refer to Precisa Customer Service If the power adaptor or its cable is damaged the balance must immediately be disconnected from the electricity supply pull out the power adaptor The balance may only be operated with a power adaptor in perfect condition DANGER Warning of a p...

Страница 10: ...balance must not be operated in an area subject to explosion risks Take care when weighing liquids that no liquid is spilt into the inside of the balance or into connections on the rear of the equipment or the power adaptor If liquid is spilt on the balance the latter must immediately be unplugged from the mains electricity supply pull out power adaptor The balance may be operated after it has fir...

Страница 11: ... 1 the scale pan holder 4 and the pan 5 which depending on the model may be square fig 2 1 right or round fig 2 1 left Depending on the model see Chapter 2 3 Data and parameters the balance additionally has a simple windshield fig 2 1 right or a windshield with sliding doors fig 2 1 left 2 and or a protective ring 3 Fig 2 1 Your balance ...

Страница 12: ...the XB series of balances Precisa also offers balances protected against water and dust to the international 1P65 standard Virtually all models of the XB Series can be delivered in calibratable design to EG OIML specification The most important basic production features of the Precisa XB Series include Anti theft encoding with four figure numerical code Multi stage password protection for the prog...

Страница 13: ...e device for under floor weighings 2 Nameplate 3 Adjustable feet for levelling 4 Weighing pan 5 Multifunction display 6 5 key control panel 7 Serial No plate 8 Connecting socket for power adaptor 9 RS232 Interface 10 Bubble level 11 Eyelet for attaching a safety chain ...

Страница 14: ...ns must be obser ved 2 2 2 Conformity The balance has been manufactured and tested in accordance with the standards and recommendations set out in the enclosed certificate of conformity The power adaptor produced for the operation of the balance and intended exclusively for this application complies with the electrical protection class II 2 2 3 Labelling Label Location of label Manufacturer s name...

Страница 15: ...specification e g A analytical balance M MilIigramme balance the number before it corresponding in each case to the maximum allowable load in grammes The allowable weighing range the calibration value and the readability of the balance are printed on the casing see Tab 2 1 List of appliance labels and are therefore not presented here 6 7 XB 2200 C Max 2200 g Min 0 5 g e 0 1g d 0 01g 2HA EI 1 2 3 4...

Страница 16: ...dust and water splash protection to IP65 Precisa XB D e g XB 3200 D and XB G e g XB 10200 G square weighing pan 200 x 200 mm Fluorescent or LCD Display also available as Floating Range version also available with dust and water splash protection to IP65 The following applies to all Precisa XB balances Mains connection 115 or 230V 15 20 50 to 60Hz Power consumption without peripheral appliances 6 0...

Страница 17: ...packing Check the balance immediately after unpacking for externally visible damage If you should find transport damages please inform your Precisa Services representative immediately If the balance is not to be used immediately after purchase but only at a later time it should be stored in a dry place where fluctuations in temperature are as low as possible see Chapter 12 Transport storage and di...

Страница 18: ...delivery broken down by groups Component delivered present yes no Balance Scale pan support Weighing pan Power adaptor Protection cover for the display Windshield BASIC group A Windshield simple group M Protective ring groups A M and C Operating instructions Guarantee card Conformity note Tab 3 1 Check list for the scope of delivery ...

Страница 19: ...lance so that the allowable ambient conditions see Chapter 2 3 1 Technical data are maintained and in addition so that the following prerequisites are met Put the balance on a solid firm and preferably vibration proof horizontal base Make sure that the balance cannot be shaken or knocked over Protect from direct solar radiation Avoid drafts and excessive temperature fluctuations NOTE A screwdriver...

Страница 20: ...ing The two screw feet must be adjusted so that the air bubble is preci sely in the centre of the sight glass of the bubble level see Fig 3 2 Correct levelling with the aid of the bubble level Fig 3 2 Correct levelling with the aid of the bubble level DANGER The balance may only be operated with the power adaptor supplied Check before connecting the power adaptor to the mains supply that the opera...

Страница 21: ...lance can itself determine the size of the calibration weight which enables an exact calibration with different size weights in 10 g 50 g 100 g and 500 g steps depending on implementation 3 8 Dual Range and Floating Range balances With the Precisa Dual Range balances weighing is always first carried out in the small more exact range Only when the small range is exceeded will the balance switch aut...

Страница 22: ... the scale pan support and turn the balan ce around Push the metal cover 1 on the floor of the balance to one side Hang a small hook available as an accessory see Chapter 13 Ac cessories into the aperture 3 of the now visible metal casting 2 Place the balance over an opening Replace the scale pan support and the scale pan Level the balance see Chapter 3 6 Levelling the balance Switch on the balanc...

Страница 23: ...N Take care that the hooks used for the under floor weighing are stable enough to hold the goods which you wish to weigh NOTE Take care that no dirt or moisture can get into the balance with the scale pan carrier removed After completing the under floor weighing the opening in the floor of the balance must be closed again dust protection kg 1 2 3 2 ...

Страница 24: ... mode is activated the balance automatically switches to Standby some time after the last weighing or key operation current saving function The delay before switching to Standby is defined in the configuration menu see Chapter 5 7 Weighing mode Press any button or put on a weight in order to switch the balance from the Standby mode back to the weighing mode again 4 3 Setting and storing the config...

Страница 25: ...your specific needs In the application menu you define a working program which is suited to the specific weighing problem 4 5 Activating the two main menus 4 5 1 Activating the configuration menu Press ON OFF to switch on the balance Continue to hold down the MODE button during the start up process approximately 10 seconds until the currently set language appears Now you can change the configurati...

Страница 26: ...which the parameters for the different function programs of the balance are defined Moving in the menu path MODE short In path to the right or start input MODE long In path downwards PRINT button In path to the left in main path exit menu Input MODE short Change value or position MODE long Change one position to the right or terminate input PRINT button Terminate input ...

Страница 27: ...ting 25 4 6 1 Control panel Four of the five keys of the Multifunctional Control Panel serve multi ple functions functions for the weighing mode and for the programming mode respectively Fig 4 1 The Five key Control Panel swissmade ...

Страница 28: ... and off MODE Calling up the configuration menu and the application menu Switches between the Basic program and the chosen application T Initiate Tare Function and or Calibration Function REF Run application see Chapter 6 Working with the application menu PRINT Start print function Exit the menu Tab 4 1 Key functions NOTE For the operation of the T and PRINT see Chapter 7 Special operating keys ON...

Страница 29: ...ance can be protected by a freely selectable four figure password against unintentional changes With password protection deactivated any operator can change the balance configuration and application menus With standard password protection activated only the configura tion menu is protected against unintentional changes With high password protection activated both the configuration menu and the app...

Страница 30: ...ode After seven consecutive incorrect entries the display will read NO ACCESS CALL SERVICE In this case only Precisa Service can unblock the balance again To activate the anti theft encoding and to alter the code programmed in the works to one which you have chosen yourself see Chapter 5 11 Anti theft encoding NOTE The password protection is deactivated at the works The pre programed password set ...

Страница 31: ...for balances with LCD dis play UNIT 1 Unit in which the results of the weighing are dis played SET DATA PRINT Print formats Type of values to be printed indivi dual values continuous printout time or load change dependent values date time user etc SET CALIBRATION Method of calibration SET WEIGHING MODE Stability mode Quality of the balance location Auto Standby mode zero correction tare method rap...

Страница 32: ...e menu see Chapter 4 5 Activating the two main me nus und Chapter 4 6 How the menu control operates In order to alter the language proceed as follows Switch off balance Switch on balance and depress MODE until the currently activated language is shown Release MODE Press MODE repeatedly until the language you require is shown Press PRINT in order to confirm the selection Press PRINT in order to qui...

Страница 33: ...use of the corresponding weight range Tab 5 2 Possible weight units and conversion factors UNIT 1 UNIT 1 g kg Bht Gramme Kilogramme Baht Display Weight unit Conversion to grammes g Gramme mg Milligramme 0 001 g kg Kilogramme 1000 g GN Grain 0 06479891 g dwt Pennyweight 1 555174 g ozt Troy ounce 31 10347 g oz Ounce 28 34952 g Lb Pound 453 59237 g ct Carat 0 2 g C M Carat Metric 0 2 g tLH Tael Hong ...

Страница 34: ...this product name is stored SET DATA PRINT AUTO START ON OFF Start print automatically on switching on off MODE UNSTABLE Individual print each value MODE STABLE Individual print stable value MODE LOADCHANGE Print after load changes MODE CONTINUOUS Continuous print after every lntegration time MODE TIMEBASE Continuous print with time basis TIMEBASE 2 0 Time basis in seconds freely selectable SET PR...

Страница 35: ...nd Chapter 4 6 How the menu control operates For the calibration of the balance see Chapter 3 7 Calibration of the balance and Chapter 14 2 Notes on calibration SET CALI BRATION MODE OFF Closed MODE EXTERNAL External MODE EXT DEF External with user defi ned weight DEF n nnn g MODE INTERNAL with internal weight MODE AUTO Automatic AUTOCAL DEF 0 000 g Calibration weight for EXT DEF mode AUTOCAL TIME...

Страница 36: ... automatically into the energy saving mode SET WEIGH ING MODE FLOATINGDISPLAY 0 04 FLOATINGDISPLAY 0 08 FLOATINGDISPLAY 0 16 FLOATINGDISPLAY 0 32 Input lntegration time in seconds STABILITY LOW STABILITY MEDIUM STABILITY HIGH Setting the Stability control instability of the balance location AUTO STANDBY OFF AUTO STANDBY 5 MIN AUTO STANDBY 10 MIN AUTO STANDBY 30 MIN Auto Standby not acti ve or acti...

Страница 37: ...232 V24 interface of the balance is matched to the interface of a peripheral device see Chapter 8 Data transfer to peripheral devices SET INTERFACE BAUDRATE 300 BAUDRATE 600 BAUDRATE 1200 BAUDRATE 2400 BAUDRATE 4800 BAUDRATE 9600 BAUDRATE 19200 Select baud rate PARITY 7 EVEN 1STOP PARITY 7 ODD 1STOP PARITY 7 NO 2STOP PARITY 8 NO 1STOP Select parity HANDSHAKE NO HANDSHAKE XON XOFF HANDSHAKE HARDWAR...

Страница 38: ...ection of the menus and Chapter 10 1 4 Activation of password protection for further information on password protection SET DATE AND TIME DATE DD MM YY TIME HH MM SS FORMAT STANDARD US Set date and time NOTE These settings are only available if the clock is installed The date and time display continues in the event of a power failure If this is not the case the balance backup battery is exhausted ...

Страница 39: ...ncoding is activated a four figure code must be entered after every interruption of the power supply in order to release the balance for use For further information on the anti theft encoding see Chapter 4 8 Anti theft encoding To activate the anti theft encoding proceed as described for pass word protection THEFTCODE THEFTCODE THEFT PROTECTION ON OFF Switch encoding on off NEW CODE Enter new code...

Страница 40: ...menu and adapted to the user s needs Printing conventions The settings for the sub menus programmed in works are printed in bold in these operating instructions For a greater clarity only that part of the menu tree which corresponds to this application is shown with each application description You will find the complete menu tree for the application menu in Chapter 16 Application menu tree Explan...

Страница 41: ...ol operates 6 3 Changing application weighing mode To activate an application it must first be selected in the application menu see Chapter 6 2 Selecting a working program Press the MODE button The display will show either the name of the selected application UNITS COUNT or BALANCING for switching back to weighing mode If SET APP OFF appears immediately no application has yet been selected Release...

Страница 42: ...mode Press the REF button until the desired unit appears on the display Release the REF button SETUP APPLICATION UNITS UNIT 2 kg kilogramme UNIT 2 mg milligramme UNIT 2 UNIT 2 OFF not active UNIT 3 GN Grain UNIT 3 UNIT 3 OFF not active UNIT 4 C M Metric carat UNIT 4 UNIT 4 OFF not active NOTE For basic operation Unit 1 is defined in the configuration menu standard unit for all weighings if the wor...

Страница 43: ...s for the regulation of the reference weight Operation Start the application see Chapter 6 3 Changing application weig hing mode Place a number of objects on the weighing pan Press the REF button until the reference quantity appears on the display If the reference quantity is correct release the REF button Otherwise hold down the REF button The pre defined values will appear 1 10 25 50 As soon as ...

Страница 44: ...fined reference weight Operation Start the application see Chapter 6 3 Changing application weig hing mode Place the reference weight on the weighing pan Press the REF button until REFERENCE 100 appears on the display Release the REF button 6 7 Application Net Total There is no Setup menu for this application With the aid of the NET TOTAL application you can add individual weighings where the bala...

Страница 45: ...er 6 3 Changing application weigh ing mode Place something on the weighing pan Sto n Take stable value and add to the sum of the components Tare balance Short confirmation with display StorEd n rESEt Reset InFo Display of Total weight TOT Residual capacity R C Individual components n Exit info with MODE button ...

Страница 46: ... 6 Working with the application menu 44 Press the REF button until the desired function appears on the display Release the REF button ...

Страница 47: ...d until CALIBRATION is displayed Release T The balance carries out a calibration in accordance with the set tings in the configuration menu cf Chapter 5 6 Calibration functions and reports these by means of a printout Carries out a calibration test Ensure that the balance is in the weighing mode Keep T pressed until TEST CALIBRATION is displayed Release T The balance carries out a calibration test...

Страница 48: ...mode Keep PRINT pressed until RESET PROD COUNTER is displayed Release PRINT The product counter will be reset to 1 Print out a balance status Ensure that the balance is in the weighing mode Keep PRINT pressed until PRINT STATUS is displayed Release PRINT The balance status will be printed out Print out the application setup Ensure that the balance is in the weighing mode Keep PRINT pressed until P...

Страница 49: ...he handshake is set to NO none in the works It can be set to software handshake XON XOFF or to hardware handshake Baud rate Possible baud rates 300 600 1200 2400 4800 9600 or 19200 baud Parity Possible parity 7 even 1 stop 7 odd 1 stop 7 No 2 stop or 8 No 1 stop Parity SB 1 2 3 4 5 6 7 8 SP 7 even 1 SB 1 DA 2 DA 3 DA 4 DA 5 DA 6 DA 7 DA PB SP 7 odd 1 SB 1 DA 2 DA 3 DA 4 DA 5 DA 6 DA 7 DA PB SP 7 n...

Страница 50: ...t Balance RJ 45 D25 D9 Peripheral device RS 232 out 2 3 2 RS 232 in RS 232 in 6 2 3 RS 232 out GND 5 7 5 GND Balance RJ 45 D25 D9 Peripheral device RS 232 out 2 3 2 RS 232 in RS 232 in 6 2 3 RS 232 out GND 5 7 5 GND CTS 3 20 4 DTR DTR 7 5 8 CTS Balance RJ 45 Remark n c 1 Not connected RS 232 out 2 Out V24 CTS 3 In V24 VDC 4 Out 9 16V GND 5 0V RS 232 in 6 In V24 DTR 7 Out V24 EXTBUS 8 In 5V logic 8...

Страница 51: ... D1 D0 U U U 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 B B B S D7 D6 D5 D4 D3 D2 D1 DP D0 B U CR LF B Blank space S Sign space DP Decimal Point D0 D7 Digits U Unit only when weight is stable else no unit transfered CR Carriage Return LF Line Feed NOTE Unused positions are filled with spaces The decimal point DP can be between D0 and D7 ...

Страница 52: ...Cxxxx Enter anti theft code PDT Print out date and time PRT Start printing Press Print key PST Start print status Pn ttt t Set print mode n 0 Individually print each value unstable n 1 Individually print each value stable n 2 Print after change of load n 3 Print after each integration period n 4 Print on time basis in s ttt t R k Set current weight 100 with k 0 7 decimal places k A use automatic p...

Страница 53: ...utes Seconds T ttt Tare or set tare to a specific value Uxnn Set unit x 1 4 of the balance with nn 0 g 1 mg 2 kg UxS Switch balance to unit x 1 4 ZERO Zero balance provided weight is stable and within the zero position range NOTE Each remote control command must terminate with CR LF The commands are acknowledged if required Input Description of the function executed D Five dashes will be shown DTE...

Страница 54: ... to power adaptor is interrupted Power supply has failed interruption to cur rent The power adaptor is defective OL is shown in display The weight range has been exceeded Observe information on the maximum weight range UL is shown in display The weight range is below the range of the ba lance Scale pan or scale pan holder missing The weight display fluctuates continuously The draft is too strong a...

Страница 55: ... or only dashes The stability control Balance functions is set too sensitively The time selected for FLOATINGDISPLAY is unsatisfactory Configuration menu cannot be changed The password lock is activated in the configu ration menu The display flashes conti nuously du ring calibration The balance location is not quiet enough Inter rupt calibration with ON OFF and relocate the balance in a better pos...

Страница 56: ... hold down until the currently set language appears Release the MODE button or LANGUE FRANCAISE or LANGUAGE ENGLISH LANGUAGE ENGLISH Press MODE repeatedly until the desired language is displayed the display will flash Hold down MODE until entry is complete display ceases to flash or press PRINT Press PRINT to exit menu unit 1 9 Press MODE and hold down until the currently set unit appears Release ...

Страница 57: ...INT Press PRINT again to exit menu SET DATA PRINT Press MODE and hold down until SET DATA PRINT appears Release the MODE button AUTO START OFF Press MODE briefly AUTO START will appear AUTO START ON Press MODE repeatedly until the desired setting is displayed ON or OFF will flash mODE STABLE Press MODE and hold down the button until the next print function appears MODE Release the MODE button MODE...

Страница 58: ...0 Press MODE until the second digit flashes Enter the second digit by briefly pressing MODE repeatedly PASSWORd 7914 In the same way enter the third and fourth digit of the password Hold down MODE until input is complete no digit flashing DATA PROTECTION OFF Press MODE briefly DATA PROTECTION will appear Press MODE repeatedly until the desired data protection is displayed the setting will flash ne...

Страница 59: ...Release the MODE button SET APP count Press MODE repeatedly until SET APP COUNT is displayed the application will flash Press PRINT to confirm the input SETup APPlication Press MODE until SETUP AP PLICATION appears Release the MODE button Reference 10 pcs Press MODE briefly the cur rently set reference quantity ap pears Release the MODE button Reference 010 pcs Press MODE briefly The first digit o...

Страница 60: ...set number of objects on the weighing pan The reference weight for the count will be defined by briefly depressing the REF button Reference 025 pcs In the same way enter the third digit of the reference quantity Hold down MODE until input is complete no digit flashing Press PRINT twice to exit menu ...

Страница 61: ...y years and will extend the lifespan of the balance DANGER For maintenance work the balance must be separated from the power supply remove power adaptor plug from socket Also ensure that the balance cannot be reconnected to the power supply during the work by a third party Take care during cleaning that no liquid penetrates into the appliance If liquid is spilt on the balance the latter must immed...

Страница 62: ...the balance If you would like to take the balance out of service for an extended period disconnect it from the electricity supply clean it thoroughly see Chapter 11 Maintenance and servicing and store it in a place which fulfils the following conditions No violent shaking no vibrations No large temperature fluctuations No direct solar radiation No moisture NOTE The balance should preferably be dis...

Страница 63: ...s Parts counting with stored values and part numbers 350 8511 001 Finished pack control to FPVO 350 8511 002 Signal lamp with 3 bulbs green yellow red 350 8510 Input Output module 6 TTL inputs 8 Relay outputs 350 8509 Multiplexer for up to 7 balances RS232 350 8513 Data cable RJ45 RJ45 0 75 m 350 8525 Data cable RJ45 RJ45 1 5 m 350 8520 Data cable RJ45 RJ45 3 m 350 8521 Data cable RJ45 DB9 female ...

Страница 64: ...ory Article number Density determination kit 350 8515 Windshield for M and C balances Heighth 180 mm Heighth 260 mm 350 8518 350 8519 Dust cover for display 350 4096 Night cover 350 4097 Hooks for under floor weighing 350 8527 ...

Страница 65: ...ility control depends on the quality of the balance location and must be correctly chosen in order to obtain optimal reproducible results Choose HIGH STABILITY at an optimal balance location MEDIUM STABILITY at a good balance location or LOW STABILITY at a critical balance location Optimal balance location FLOATINGDISPLAY 0 04 or FLOATINGDISPLAY 0 08 Good balance location FLOATINGDISPLAY 0 16 Crit...

Страница 66: ...ive a remote control command via the interface 14 1 4 Set Weighing mode Auto Zero If the automatic zero correction Auto Zero is activated the balance always gives a stable zero e g even with room temperature fluctua tions 14 2 Notes on calibration The calibration of the balance is fixed in the configuration menu see Chapter 3 7 Calibration of the balance and Chapter 5 6 Calibration functions Possi...

Страница 67: ...ALIBRA TION MODE EXT DEF must be selected in the configuration menu see Chapter 5 6 Calibration functions Then the effective value of the calibration weight DEF n nnn g must be entered with up to tenfold precision compared with the balance Switch to BALANCING with the MODE key cALIBRation Press T until CALIBRATION appears 0000 9 The balance carries out a Zero measurement 0 000 g is shown flashing ...

Страница 68: ... Press T until CALIBRATION is shown The calibration is finished after a certain period of time Switch to BALANCING with the MODE key cALIBRation Press T until CALIBRATION appears 0000 9 The balance carries out a Zero measurement 0000 g is shown flashing 372 9 After the zero measurement the display flashes with the previous ly entered calibration weight Place the calibration weight on the pan 372 9...

Страница 69: ...the automatic calibration is as determined in the configu ration menu under SET CALIBRATION AUTOCAL TIME n h e g 6 h for 06 00 o clock in the morning NOTE For the automatic calibration by time and by time temp the date and time of the balance must first be correctly set see Chapter 5 9 Date and time The calibration can also be effected manually any time while auto calibration is activated The auto...

Страница 70: ...on switching on LANGUAGE LANGUAGE ENGLISH SPRACHE DEUTSCH LANGUE FRANCAISE CONTRAST CONTRAST 6 UNIT 1 UNIT 1 g mg kg GN dwt ozt oz Lb ct C M tlH tlM tlT mo t Bht SET DATA PRINT AUTO START ON OFF MODE UNSTABLE MODE STABLE MODE LOADCHANGE MODE CONTINUOUS MODE TIMEBASE TIMEBASE 2 0 ...

Страница 71: ...ERATOR ttt SET CALI BRATION MODE OFF MODE EXTERNAL MODE EXT DEF MODE INTERNAL MODE AUTO DEF 0 000 g AUTOCAL TIME TEMP AUTOCAL TEMPERATURE AUTOCAL TIME AUTOCAL TIME 6 h SET WEIGH ING MODE FLOATINGDISPLAY 0 04 FLOATINGDISPLAY 0 08 FLOATINGDISPLAY 0 16 FLOATINGDISPLAY 0 32 STABILITY LOW STABILITY MEDIUM STABILITY HIGH AUTO STANDBY OFF AUTO STANDBY 5 MIN AUTO STANDBY 10 MIN AUTO STANDBY 30 MIN AUTO ZE...

Страница 72: ...TY 7 EVEN 1STOP PARITY 7 ODD 1STOP PARITY 7 NO 2STOP PARITY 8 NO 1STOP HANDSHAKE NO HANDSHAKE XON XOFF HANDSHAKE HARDWARE SET DATE AND TIME DATE DD MM YY TIME HH MM SS FORMAT STANDARD US PASSWORD PASSWORD DATA PROTECTION OFF DATA PROTECTION MED DATA PROTECTION HIGH NEW PASSWORD THEFTCODE THEFTCODE THEFT PROTECTION OFF THEFT PROTECTION ON NEW CODE ...

Страница 73: ...ssed on switching on SET APP OFF UNITS COUNT PERCENT SETUP APPLICATION Division depends on the current application see Chapter 6 Wor king with the application menu AUTO START AUTO START ON OFF FACTORY CONFIG Load factory configuration USER CONFIG Load user configuration STORE CONFIG Store present configuration as user con figuration ...

Отзывы: