FASTCAM-ultima APX-i2 Hardware Manual
-9-
1.4. Precautions
Ambient Temperature
Photron FASTCAM-ultima APX-i2 has been designed to work properly in an ambient
temperature range of 0 to 35 degrees Celsius (32 to 95 degrees Fahrenheit), no
condensation.
Storage Temperature
The APX-i2 must be stored in a place with an ambient temperature range of -20 to +70
degrees Celsius (-4 to +158 degrees Fahrenheit), no condensation.
Transportation
Save the original carton the unit came in for future transportation.
Do not apply excessive shocks or vibration during transportation, storage or operation.
The image intensifier has a glass tube that is fragile and operates at a high degree of
vacuum. Excessive shocks or vibration may cause failures or malfunctions.
Do not transport the unit under ambient temperature of below -20 degrees Celsius (-4
degrees Fahrenheit) or above 70 degrees Celsius (+158 degrees Fahrenheit).
No condensation should be allowed during transportation, storage or operation.
FEDERAL COMMUNICATIONS COMMISSION STATEMENTS
WARNING: This equipment generates, uses and can radiate radio frequency energy
and if not installed and used in accordance with the instruction manual, may cause
interference to radio communications. It has been tested and found to comply with the
limits for a Class computing device pursuant to Subpart A of Part 15 of the FCC Rules
and VDE 0871 Class B which are designed to provide reasonable protection against
such interference when operated in a commercial environment. Operation of this
equipment in a residential area is likely to cause interference in which case the user at
his own expense will be required to take whatever measures may be required to correct
the interference.
This device complies with Part 15 of the FCC Rules and VDE 0871. Operation is
subject to the following two conditions: (1) this equipment may cause harmful
interference, and (2) this equipment must accept any interference received including
interference that may cause undesired operation.
WARNING
This product is grounded through the power cord. This protective ground connection is
essential for safe operation of the equipment. Avoid electrical shock by plugging the
power cord into a properly wired power outlet. A loss of the protective grounding, for
any reason, could result in electrical shock. Use the proper power cord and insure that
it is in good condition.
Do not touch or allow any object to come in contact with the input window of the image
intensifier at any time. The input window can be easily damaged and electrical shocks
(static discharge) or failures may result. Never attempt to touch an internal part of
image intensifier while it is in operation.
CAUTIONS
The Image intensifiers is an extremely sensitive optical device. When operating the
system, exercise caution to gradually increase light falling on the light-entrance window
Содержание FASTCAM ultima APX-i2
Страница 1: ...FASTCAM ultima APX i2 Hardware Manual Rev 1 04 PHOTRON LIMITED SEP 2007...
Страница 2: ...FASTCAM ultima APX i2 Hardware Manual...
Страница 6: ...FASTCAM ultima APX i2 Hardware Manual 4 Memo...
Страница 13: ...FASTCAM ultima APX i2 Hardware Manual 11 Memo...
Страница 35: ...FASTCAM ultima APX i2 Hardware Manual 33 Memo...
Страница 49: ...FASTCAM ultima APX i2 Hardware Manual 47 Memo...
Страница 88: ...FASTCAM ultima APX i2 Hardware Manual 86 TRIGGER SW IN Circuit TRIGGER TTL IN Circuit GENERAL IN Circuit...
Страница 127: ...FASTCAM ultima APX i2 Hardware Manual 125 Memo...
Страница 145: ...FASTCAM ultima APX i2 Hardware Manual 143 7 2 3 Keypad millimeters 139 5 81 5 81 5 30 3 37 8...