G 03895 - Edition 03 - February 10
1/1
B 100
Emergency stop plug connection
1
2
4
3
This connector allows the
control or remoting of the
emergency stop function
from the front panel of
the unit.
This connector is located at
the rear of the pump.
“J1“ connector
location
Cover plug wiring
In local mode, the pump can run only if the cover
plug (delivered with the pump) is fitted on “J1“
connector.
When the emergency stop is interfaced, if the
remote mode is not used, it is necessary to connect
the cover plug: “J14“.
AMP connector
Wiring viewed under
soldered side
All dry contacts are rated 50V - 1A.
The input is considered activated when the pins are linked.
Function
2-3
Contact opened, emergency stop command is activated.
Function
1-4
This contact opens when emergency stop is activated:
Contact opened, emergency stop command is activated.
The output contact connected by the customer allows to control the
equipment emergency stop from the pump emergency stop button.
Control contacts
(input)
Available output:
“Emergency stop“
state
When units containing control circuits are equipped with dry contact outputs,
it is the responsability of the customer to use these outputs in compliance
with safety extra low voltage installation and security standards : it concerns
J6, J7 and J14 connectors.
J14
J1
adixen Vacuum Products - A3H Series Operating instructions