4/4
C 80
Use the bellows heater
G 04796 - Edition 01 - August 12
Parameters list
Menu
Parameter Description
Parameter value
DiSP
Std
Display process value and set point accessed by pressing the and buttons
Regulated Value – set
point value 120 °C
AL
2FSH
Alarm on High value (max. temperature allowable in continuous on RSV
heaters device )
140 °C
HY
Alarm hysteresis
1
Atun
tunE Automatic
tuning
OFF
PiD
Pb
Proportional band
20
ti
Integral constant
360
td
Derivative constant
60
Hcb Cutback
high
Auto
Lcb Cutback
low
Auto
SP
SP L
Set point limits low limit
0
SP H
Set point limits high limit
140
SPrr
Set point speed limit
OFF
tmOP Timer
mode
OPt1
tmr Time
remaining
0
dwell Palier
time
OFF
Stat
Timer state
OFF
iP
FiLt Filter
time
in
seconds
1,6
OFS
Offset of process value
0
CAL.P Access
code
for
calibration
0
oP CYCH
Heat
cycle
time
1
ACCS
InSt
Unit Display
units
:
centigrade
°C
dEC.P
Decimal places in display
nnnn
Ctrl Control
type
Pid
Act
Control action : Reverse (normal action for temperature control )
rEu
Pdtr
Manual reset tracking (PD control) In auto holds manual reset value
Hold
iP
inPt
Type of temperature sensor : PT 100
rtd
CJC Cold
compensation
solding
Auto
ImP Impedance
adapter
value
in
case
of
sensor
loss
Auto
AL
AL 1
Alarm 1 - disabled
OFF
Ltch
Alarm 1 - memorised
no
bLoc
Alarm 1 - blocked
no
AL2
Alarm 2 - High alarm mode (not used )
FSH
Ltch
Alarm 2 - memorised
no
bLoc
Alarm 2 - blocked
no
AL3
Alarm 3 - disabled
OFF
AL3
Alarm 3 - memorised
no
AL3
Alarm 3 - blocked
no
SPLi
Limitation of Alarm set point by set-up limits
Con
AA
id Output
relay
rELY
Func Logic
output
diG
IA
id
Switched DCLogic output
LoG
Func Heater
output
HEAt
diGF
Alarm mode if safety sensor broken
Sbr
SEnS Output
relay
type
nor
adixen Vacuum Products - A3H Series Operating instructions