G 03912 - Edition 04 - August 12
1/4
D 10
First level of maintenance
Maintenance time
monitoring
Maintenance
frequency
When a problem occurs, the user is informed by:
the relevant fault indicator light which is located on the
front panel: yellow lit in case of warning, red lit in case of
fault (followed by a pump stop or not, depending on setting
parameters,
C 42
),
the audible buzzer warning (if preselected),
the fault contacts on the J14 remote connector at the rear of the
monitoring system which close,
the stopping of the pumping although the STOP command has
not been given,
the display of fault message on the display of the hand-held
remote control,
a message via the RS 232 or RS 485 link on a micro-computer.
Depending on the maintenance parameter programmed, when the
operating time before maintenance has expired, the screen display
alternates between operating parameters and the corresponding
warning message.
The red indicator light is lit on front panel.
This information can be read at any time using the MAINTENANCE
menu.
Perform maintenance on the relevant parts.
Maintenance times can change according to processes and
equipment used. Contact the pump manufacturer for more
information.
There is no oil change required on site. This is done during the
pump overhaul by the manufacturer’s trained personnel.
Study the safety instructions related to maintenance (
D 00
).
Oil change
adixen Vacuum Products - A3H Series Operating instructions