G 03910 - Edition 04- August 12
1/6
C 90
Use of the serial link (M4 monitoring)
Press
SET
to enter in menus.
Press the key
+
repeatedly to have
access to the SETTING menu.
Valid with the key
ENTER
.
Valid SERIAL LINK menu with the
key
ENTER
.
Access to the parameter with
+
and valid with the key
ENTER.
Modify the chosen parameter with
the keys
+
and
–
, and valid the
setting with key
ENTER.
Return to main menu with key
SET
.
Access in the
“SETTING” menu
M4 Monitoring
system parameter
configuring
The initial configuration of the serial link is as follows:
Type:
RS 232
ECHO:
ENABLED
Transmission SPEED:
9600 bauds
Parity:
NO
2 Stop bits :
DISABLED
Address:
0
Setting / Observations
Action / Observations
Display / Observations
Access to the parameter
programming by the
“Serial link” menu
Access to the parameter
to modify
DEFINITION MENU
>>> SETTING <<<
MAINTENANCE
MANAGEMENT
PURGE PROLONGED
>>> SERIAL LINK <<<
DATE TIME
TEMPERATURE UNIT
>>> TYPE RS 232 <<<
ECHO ENABLED
SPEED 9600
PARITY NO
Controlling the unit
pumping using the
serial link
The controls are used to manage information relative to the
working pump. The pumping remote controls are however
managed as a priority.
adixen Vacuum Products - A3H Series Operating instructions