THEORY
PerkinElmer Life and Analytical Sciences
205
Operational Considerations
The operating procedures for the Tri-Carb 3170TR/SL liquid scintillation analyzer
are described in the appropriate areas of this documentation. Aside from the
count rate limitations and differences in the SNC/IPA procedures, operation of the
Model 3170TR/SL analyzer is exactly the same as other Tri-Carb 3100 analyzers.
The selection of counting vials for low level counting is very important to assure
low background and high sample counting sensitivity. Low potassium glass vials
are recommended when counting low levels of
3
H or
14
C. Certain plastic vials may
also be used for large or small samples measuring
3
H in water.
It is recommended that all glass or plastic vials used for low level counting be
carefully sealed. Teflon cap liners for the screw caps of the vials improve the seal
against leakage.
Glass vials should be washed, then rinsed in a dilute ethylene-diamine-tetraacetic
acid (EDTA) solution, followed by a “dead’’ deionized water rinse and allowed to
air dry. This will help reduce background and further reduce surface contaminants
which could contaminate the detector guard.
Maximum vial dimensions are as follows: small vials must not exceed 17.8mm
diameter and 58.0mm height (including cap); large vials must not exceed 28.1mm
diameter and 63.0mm height (including cap).
Содержание QuantaSmart 1694267
Страница 1: ...QuantaSmart For The TriCarb Liquid Scintillation Analyzer Reference Manual...
Страница 9: ...v Chapter 12 Theory 197 Low Level Counting Theory 197 Alpha Beta Counting Theory 207...
Страница 10: ...vi...
Страница 44: ...CHAPTER 3 34 PerkinElmer Life and Analytical Sciences...
Страница 118: ...CHAPTER 5 108 PerkinElmer Life and Analytical Sciences...
Страница 192: ...CHAPTER 9 182 PerkinElmer Life and Analytical Sciences...
Страница 198: ...CHAPTER 10 188 PerkinElmer Life and Analytical Sciences...