CHAPTER 7
132
PerkinElmer Life and Analytical Sciences
If the instrument is IDLE (not counting):
11. Load the calibration cassette on the right-hand side of the sample changer
deck such that you can read the protocol flag.
Press the Start button to begin counting.
If the instrument is currently counting a sample:
Load the calibration cassette after the last cassette belonging to the current
protocol. When the current protocol is completed, the calibration cassette will
automatically move into the counting position. After the SNC flag is read by the
instrument, the calibration, normalization and IPA procedure begins and the
flag is automatically returned to the “non-reset” position.
To access the data generated from the IPA runs, select IPA Charts & Tables
(optional on the Tri-Carb 2900TR) from the IPA menu.