CHAPTER 9
178
PerkinElmer Life and Analytical Sciences
Shutter open.
A Sample changer error has occurred.
No action is required unless a message is printed. If this message appears on a
printout, call PerkinElmer Technical Service.
Skipped cassette with no label.
A cassette on the sample changer deck is without an ID label.
Attach an ID label to the cassette.
The instrument encountered a protocol flag for which there is no
assay associated. This protocol was not counted.
There is a cassette on the instrument’s sample changer deck with a protocol flag
to which no assay has been associated.
Remove the cassette from the sample changer deck, or associate an assay to the
protocol flag number.
The nuclide associated with this assay has an incorrect number of
Quench sets.
One or more quench sets associated with a multiple label sample nuclide has
been deselected from the assay or deleted from the library.
Re-select the appropriate quench sets for the sample nuclides. Recount quench
set, if necessary.
The nuclide associated with this assay was not found in the nuclide
library. Unable to retrieve nuclide information.
The sample nuclide defined for the assay was deleted from the library.
Re-enter the sample nuclide information in the nuclide library.
The quench standard associated with this assay’s nuclide was not
found in the Quench standard library.
A quench set that was associated with a sample nuclide in an assay has been
deleted from the library.
Re-enter the standard information in the library and recount the standards.
The standard associated with this assay has not been counted.
The standard that is selected for the assay has not been counted prior to
performing the assay.
Count the standards and restart the assay.
Содержание QuantaSmart 1694267
Страница 1: ...QuantaSmart For The TriCarb Liquid Scintillation Analyzer Reference Manual...
Страница 9: ...v Chapter 12 Theory 197 Low Level Counting Theory 197 Alpha Beta Counting Theory 207...
Страница 10: ...vi...
Страница 44: ...CHAPTER 3 34 PerkinElmer Life and Analytical Sciences...
Страница 118: ...CHAPTER 5 108 PerkinElmer Life and Analytical Sciences...
Страница 192: ...CHAPTER 9 182 PerkinElmer Life and Analytical Sciences...
Страница 198: ...CHAPTER 10 188 PerkinElmer Life and Analytical Sciences...