76 . OilExpress System User's Guide
Maintenance Schedule
Maintenance must be performed on a regular basis in order to maintain the accuracy and
precision of the OilExpress System. The schedule below outlines the proper intervals to check
or replace components of the instrument.
NOTE: For information on maintaining your spectrometer, refer to the documentation
supplied with the instrument.
Key:
C = Clean this item
- = No action necessary
R = Replace component
S = Special case
X = Perform service
Item
Daily
Weekly
6 Months
ALIDUM Cable
-
X
(p.79)
R
(p.82)
Sampling Probe
C/X
(p.77)
C/X
(p.77)
R
(p.80)
Z-Rack
-
C
(p.77)
-
Splash Guard
-
C
(p.78)
-
Transfer/Wash Port
-
C
(p.77)
-
Syringe
C
(p.77)
C
(p.77)
S
(p.80)
Syringe Seal
C
(p.77)
-
R
(p.80)
Sample Rack
C
(p.77)
Sampling Probe
C
(p.77)
-
R
(p.79)
X/Y/Z Mechanism
-
-
X
(p.78)
X/Y/Z Lubrication
-
-
S
(p.84)
Tubing
-
-
S
(p.85)