32 . OilExpress System User's Guide
Use of Flammable Solvents and Samples
WARNING
The spectrometer contains a hot source and contact with flammable
vapors may cause an explosion. When working with flammable solvents
or samples, particularly during unattended operation with flow-cells, it is
recommended that the instrument optics area should be continuously
purged with dry air or nitrogen to maintain a positive pressure and
prevent flammable vapor entering the instrument.
WARNING
DO NOT use the OilExpress System when the sample compartment lid is
fitted. Flammable vapor may seep through instrument seals and gaskets
and accumulate under the lid, where it could cause a fire or explosion
hazard.
WARNING
If flammable solvents or samples are spilled on the instrument and there
is any possibility that they have entered the interior (by coming into
contact with cover gaskets for example) then the instrument must be
switched off immediately and disconnected from the power supply. The
optics area should then be thoroughly purged with dry air or nitrogen, or
the main cover should be opened to thoroughly ventilate the optics area
before proceeding.
WARNING
Flammable solvents or samples should not be stored on or near the
instrument. Handling of such materials during preparation should be
performed in a safe area away from the instrument such as a fume
cabinet.
WARNING
Some chemicals used with this instrument may be hazardous or may
become hazardous after completion of an analysis. The responsible body
(for example, the Laboratory Manager) must take the necessary
precautions to make sure that the surrounding workplace is safe and that
the instrument operators are not exposed to hazardous levels of toxic
substances (chemical or biological) as defined in the applicable MSDS
(Material Safety Data Sheets) or OSHA (Occupational Safety and Health
Administration (United States)), ACGIH (American Conference of
Governmental Industrial Hygienists (United States)), or COSHH (Control
Of Substances Hazardous to Health (United Kingdom)) documents.
Venting for fumes and disposal of waste must be in accordance with all
national, state and local health and safety regulations and laws.
Содержание OilExpress System
Страница 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 23: ...Warnings and Safety Information ...
Страница 36: ...36 OilExpress System User s Guide ...
Страница 37: ...Unpacking and Installation ...
Страница 69: ...Background to Used Oil Analysis ...
Страница 75: ...Routine Maintenance and Troubleshooting ...
Страница 96: ...96 OilExpress System User s Guide ...
Страница 97: ...Appendices ...
Страница 102: ...102 OilExpress System User s Guide ...