Routine Maintenance and Troubleshooting . 77
Autosampler Maintenance Procedures
Flushing the Syringe Pump
1.
Do not allow the pump to run dry for more than a few cycles.
2.
Inspect the pump for leaks, and correct any problems immediately.
3.
Wipe up all spills on and around the pump immediately.
Cleaning the Syringe Seal
1.
Clean the seal with heptane.
2.
Check for cracks in the Teflon coating.
Cleaning the Sampling Probe
1.
Turn off power to the Autosampler.
2.
Clean the needle by wiping gently with a lint-free cloth dampened with heptane.
Cleaning the Sample Rack
We recommend that you clean the sample rack daily with detergent or degreasant to remove
dust and oil.
Cleaning the Injection and Wash Port
Residues from samples may be deposited in the injection/wash-port. We recommend that
you remove the lid occasionally and rinse it with a suitable detergent if necessary.
Cleaning the Z-Rack
The Z-Rack must be cleaned weekly.
NOTE: Do not use alcohol or solvents when cleaning the Z-Rack.
1.
Turn off the power to the Autosampler.
2.
With a lint-free cloth, wipe the Z-Rack thoroughly.
If necessary, use a toothbrush to remove dust or dirt from the teeth of the
Z-Rack.
CAUTION
Electrostatic discharge to the sampling probe may damage the ALIDUM.
Содержание OilExpress System
Страница 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 23: ...Warnings and Safety Information ...
Страница 36: ...36 OilExpress System User s Guide ...
Страница 37: ...Unpacking and Installation ...
Страница 69: ...Background to Used Oil Analysis ...
Страница 75: ...Routine Maintenance and Troubleshooting ...
Страница 96: ...96 OilExpress System User s Guide ...
Страница 97: ...Appendices ...
Страница 102: ...102 OilExpress System User s Guide ...