48 . OilExpress System User's Guide
6.
Push the needle tubing over the shorter end of the needle (to a minimum of 25 mm,
1 inch).
7.
Loosen the Allen screw on the insulation block. Install the insulation block on the Z-rack
with the insulation block sleeve facing down and the insulation block (black coax) cable
on the left side. Ensure that the insulation block cable is not twisted.
8.
Push the insulation block onto the Z-rack until firmly seated.
9.
With an Allen wrench, tighten the insulation block set screw until it makes contact with
the flat portion of the Z-rack. Do not overtighten the setscrew.
Connecting the tubing to the solvent bottle
1.
Thread the tubing from top to bottom through the cap of the wash bottle as shown in
Figure 23 Connecting the solvent tubing
2.
Fill the solvent bottle with solvent.
3.
Screw the cap on to the solvent bottle securely.
4.
Position the bottle safely, behind the instrumentation at a location where it will not be
accidentally knocked.
Solvent bottle lid
Solvent bottle