Unpacking and Installation . 65
15.
Right-mouse click within the worktable area of the Autosampler Utilities dialog and from
the menu select
Save Worktable
.
NOTE: The worktable changes are not automatically saved. If you do not select Save
Worktable, the changes will be lost.
NOTE: The procedure described above must be repeated for all worktables used by the
OilExpress software as anchor points are not applied universally. Follow steps 16 to
20 to calibrate all additional worktables.
16.
Right-mouse click within the worktable area of the Autosampler Utilities dialog and from
the menu select
Load Existing Worktable
.
17.
Select the required worktable,
PerkinElmer BOTTLES
or
PerkinElmer VIALS
.
18.
19.
Enter the values you previously recorded in your laboratory notebook into the
X=
and
Y=
fields of the User-defined Rack Properties dialog.