64 . OilExpress System User's Guide
The anchor point depends on the version of your wash station. Examine the
injection/wash port fitted, then use the appropriate needle position anchor point, as
shown in Figure 25.
Figure 25 Injection wash port anchor positions
12.
Once the needle has been aligned with the anchor point, click the Anchor Point Insert
button .
The new numbers appear in the
X=
and
Y=
fields.
13.
Record the
Anchor Point X
and
Y
values in your laboratory notebook as the numbers
can be used in the calibration of other worktables.
14.
Click
OK
.
Needle position
anchor point
Needle position
anchor point
Needle position
anchor point
10 mm
v1
v2
v3