66 . OilExpress System User's Guide
20.
21.
If the syringes have not yet been installed, complete the hardware installation on
page 50.
Calibrating the Pathlength of the Flowcell
The OilExpress software calculates the cell pathlength by using a Partial Least Squares multi-
component calibration on heptane. A default calibration file is provided and stored as
c:\quant\pls1\methods\cellpath.md, where c:\ is the drive to which the OilExpress software
has been installed. Heptane must be used as the solvent within the OilExpress System for
accurate cell calibration.
1.
Log in to OilExpress.
2.
Select
System Settings > Instrument Settings > System Checks Setup
.
The
Check Scheduling Properties
dialog is displayed.
3.
Select
Pathlength
and click
OK
.
4.
Click the System Checks thumbnail
on the left-hand panel in the main
OilExpress dialog.
The System Checks dialog is displayed.