Unpacking and Installation . 49
Installing the Flowcell in the Sample Shuttle
WARNING
DO NOT use the OilExpress System when the sample compartment lid is
fitted. Flammable vapor may seep through instrument seals and gaskets
and accumulate under the lid, where it could cause a fire or explosion
hazard.
1.
Unpack the flowcell.
2.
Slide the flowcell into the sample shuttle (Figure 24).
3.
Make sure that the tubing cannot obstruct the optical beampath.
Figure 24 Sample shuttle accessory and flowcell fitted
Waste Container
We recommend that you use a waste container that is larger than the solvent bottle. Then, if
you empty the waste container whenever you refill the solvent bottle, the waste container
will not overflow.
The waste drains into the container by gravity, so the container should stand on a surface
that is at least 90 cm below the bench on which the OilExpress System is located.
Fitting waste tubing
1.
Attach the waste tubing to the hose connector at the rear of the transfer/wash port.
2.
Insert the waste tubing into the hole on the rear right-hand side of the base plate of the
Autosampler.
3.
Adjust the routing of the tubing so that it is not kinked.
4.
Place the other end of the waste tubing in the waste container.
Ensure that the end will not be immersed in the waste.
CAUTION
Make sure that the ends of the tubing do not become immersed in the
waste or the operation of the analyzer will be impaired.
Содержание OilExpress System
Страница 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 23: ...Warnings and Safety Information ...
Страница 36: ...36 OilExpress System User s Guide ...
Страница 37: ...Unpacking and Installation ...
Страница 69: ...Background to Used Oil Analysis ...
Страница 75: ...Routine Maintenance and Troubleshooting ...
Страница 96: ...96 OilExpress System User s Guide ...
Страница 97: ...Appendices ...
Страница 102: ...102 OilExpress System User s Guide ...