Routine Maintenance and Troubleshooting . 83
6.
Push the insulation block cable through the back end of the arm until the gold
connector is visible.
G
o l d C
o a x C
o n n e c t o r
Figure 33 Disconnecting the insulation block cable
7.
Disconnect the insulation block cable from the connector.
8.
Carefully pull the insulation block cable back out through the back of the arm.
9.
Remove the new insulation block cable and the plastic cable strap from the packaging.
10.
Feed the gold connector of the insulation block cable through the back of the arm.
11.
Attach the gold connector to the matching connector on the ALIDUM cable. Be sure that
the connectors are firmly seated.
12.
Adjust the ALIDUM cable until all of the slack is removed from behind the arm.
13.
Fasten the insulation block cable to the arm using the new plastic cable binder.
14.
Reinstall the sampling probe.
15.
Replace the top cover and retaining screws.
CAUTION
Make sure the insulation block cable is not twisted or kinked.
Содержание OilExpress System
Страница 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 23: ...Warnings and Safety Information ...
Страница 36: ...36 OilExpress System User s Guide ...
Страница 37: ...Unpacking and Installation ...
Страница 69: ...Background to Used Oil Analysis ...
Страница 75: ...Routine Maintenance and Troubleshooting ...
Страница 96: ...96 OilExpress System User s Guide ...
Страница 97: ...Appendices ...
Страница 102: ...102 OilExpress System User s Guide ...