Setting Output Processing of Sensing Information
Setting Output Processing of Sensing Information
Set how sensing information is to be processed for outputting the required values.
Setting Scaling
Change the display scale when you want to display a value different on the main display to
the actual measured value.
Place an actual sensing object in position for measurement.
There are three setting modes: “manual setting of correction values,” and “one-point
scaling” and “two-point scaling” that automatically set the correction values of a placed
sensing object.
The settings listed below return to the default settings when scaling is set. Set these items after scaling
settings have been completed.
• Zero Reset
■
Setting One-point Scaling
Measurement is performed at one position and offset values are set for that measure-
ment.
The offset and increment/decrement inversion (display inversion) can be set.
Example: Displaying the height of the sensing object
80
80
90
Sensing
object
Display value
(mm)
Actual
measured
value
(mm)
Actual measured value
80 mm
Sensor Head
Target display value
90 mm