Se
c
tion 3
SE
T
U
P
Section 3
Setting the Sensing Conditions
3-13
ZS-L
User’s Manual
Setting the Head Installation
Set how the Sensor Head is installed.
FUN Mode-[SENSING]-[SETTING]
Setting the Emitted Light Intensity
Set the amount of light emitted from the Sensor Head to match the state of the workpiece
surface.
The response may slow down if workpieces having differing reflection factors such as black-and-white
workpieces are measured with the FUN mode set to [AUTO]. In this case, narrow the adjustment range
by setting [RANGE]. If this does not increase the response speed to keep up with measurement, select
[FIXED].
FUN Mode-[SENSING]-[LASER]
Setting
Description
DIFFUSE
Select this item when the Sensor Head is installed for diffuse reflection sensing.
REGULAR
Select this item when the Sensor Head is installed for regular reflection sensing.
Setting
Description
AUTO
Automatically adjusts the emitted light amount according to the reflection
factor of the workpiece. Note that the response time varies with each mea-
surement. (default value)
RANGE
UPPER
LOWER
Sets the adjustment range for the AUTO setting.
If the response speed cannot keep up with measurement at the default
value, narrow the range.
Range: 0.1 to 80% (default: LOWER 0.1%, UPPER 80%)
Level
Sensitivity
Color of workpiece
0.1%
·
·
80%
Low
High
Bright
Dark
FIXED
Sets the emitted light amount to a fixed value.
For reference levels, see [RANGE].
Range: 0.1% to 80%.
ZS-LD20T/20ST/40T
ZS-LD50/50S/80/130/200/350S
ZS-LD20T/20ST/40T
ZS-LD50/50S/80/130/200
Содержание ZS-L -
Страница 1: ...USER S MANUAL Smart Sensor ZS L Series Ver 2 0 Cat No Z208 E1 02 ...
Страница 52: ...2 24 Section 2 INSTALLATION CONNECTION ZS L User s Manual Section 2 SmartMonitor ZS ...
Страница 90: ...3 38 Section 3 SETUP ZS L User s Manual Section 3 Setting the System Environment MEMO ...
Страница 156: ...6 46 Section 6 APPENDIX ZS L User s Manual Section 6 INDEX ...
Страница 158: ......