Section 3
Setting the Filter Function
3-19
ZS-L
User’s Manual
Setting Average
The average of the measured values obtained based on the preset number of samples
can be output. Select this setting when you want to ignore sudden changes in measured
values.
FUN Mode-[FILTER]-[AVERAGE]
Setting Differentiation
Use the differentiation function to detect only sudden changes in the measured values that
occur during very short periods of time.
The differentiation function detects changes between the present value and the measured
value that is in effect just before the comparing pitch.The coefficient of this comparing pitch
is defined as the differentiation cycle.
Example: Differentiation cycle=10 ms
FUN Mode-[FILTER]-[DIFF]
Setting
Description
1,2,4,8,16,32,64,128,256,512,
1024,2048,4096
Sets the average count. (default value: 128)
Setting
Description
OFF
The differentiation function is not used. (default value)
ON
Sets the cycle (ms) in which to perform differentiation.
Present value
Sampling
Comparison interval 10 ms
Display value
(differentiation result)
Sampling cycle
(measurement cycle)