FZ3 Processing Items List Manual
Actual coordinate
Set value
[Factory default]
Point coordinate X, Y
0.0000 to 9999.9999
[Point you tapped in the window]
Actual coordinate X, Y
-99999.9999 to 99999.9999
[0]
4.
Subsequently set up in the same manner.
5.
Tap [Generate calibration parameters].
The calibration parameters will be generated.
●
If this succeeds,
the Calibration parameter status changes from "The parameter is not generated" to "The
parameter is the latest".
●
If this fails,
"Failed to generate parameters" is displayed.
Setting Calibration through Sampling Measurement (Sampling)
This is a method for setting calibration based on measurement results.
Calibration parameters are calculated automatically when a registered model is searched and the actual
coordinates for that position entered.
1.
In the "Calibration setting" area, select "Sampling".