70
Trigger Inputs and Output Format
Section 4-3
Note
1.
The data is output in 3 digits.
2.
The data is output in 4 digits.
3.
The data is output in 6 digits.
4.
The data is output in 5 digits.
Write System Data (@SW) 1:1 Connection
Writes the specified system data to the specified parameter.
Send
Set the desired parameter number (1 to 34) to and setting. Refer to
Read System Data (@SR) for a table showing details on the system
data settings.
page 68
Receive
“OK” indicates that the command was executed correctly:
“ER” indicates that the command was not executed correctly:
35
3964(R)
Character delay time
10 to 655350 in 10-ms units.
36
Acknowledge delay
time
10 to 655350 in 10-ms units.
37
Setup attribute count
1 to 255
38
Transfer attribute
count
1 to 255
39
Master/Slave
0: Master, 1: Slave
Number
System data
@ S
<Footer>
W
<Parameter no.>
<Setting>
O K <Footer>
<Header>
E R <Footer>
<Header>
Содержание V530-R150E-3 -
Страница 1: ...Cat No Z155 E1 01 2 Dimensional Code Reader Fixed Type V530 R150E 3 V530 R150EP 3 OPERATION MANUAL...
Страница 2: ...V530 R150E 3 V530 R150EP 3 2 Dimensional Code Reader Fixed Type Operation Manual Produced August 2001...
Страница 7: ......
Страница 9: ......
Страница 17: ......
Страница 27: ......
Страница 47: ......
Страница 63: ......
Страница 224: ...207 System Section 6 4 2 Set each item 3 Select End before leaving this screen...
Страница 226: ...209 System Section 6 4 2 Select the error method 3 Select End to leave this screen...
Страница 247: ...230 2 Dimensional Code Reader Section 8 1 8 1 2 Dimensional Code Reader V530 R150 Four 4 5 dia 165 min Unit mm...
Страница 263: ......
Страница 279: ......
Страница 281: ......
Страница 283: ......
Страница 289: ......