194
System
Section 6-4
6-4-2
Communications Method
Sets communications methods when communicating to external
devices via terminal blocks or RS-232C.
• When communicating in normal format via RS-232C:
Set the communications method for RS-232C and Normal.
• When communicating with host link via RS-232C:
Set the communication method for RS-232C and Host link.
• When inputting trigger or outputting reading results (OK/NG) to
the OK/NG terminals on terminal blocks:
Set the communications method for Terminal block.
RS-232C
1,2,3...
1.
Select 2400, 4800, 9600, 19200, or 38400 (bps).
(Default setting: 38400)
2.
Select 7 or 8 (bits). (Default setting: 8)
3.
Select None, Odd, or Even. (Default setting: None)
4.
Select 1 or 2 (bits). (Default setting:1)
5.
Select None, STX, or ESC. (Default setting: None)
6.
Select CR, LF, or CR+LF, ETX. (Default setting: CR)
7.
Select OFF or ON. (Default setting: OFF)
8.
Select Normal, Host link, 3964, or 3964R (Default setting: Nor-
mal)
Содержание V530-R150E-3 -
Страница 1: ...Cat No Z155 E1 01 2 Dimensional Code Reader Fixed Type V530 R150E 3 V530 R150EP 3 OPERATION MANUAL...
Страница 2: ...V530 R150E 3 V530 R150EP 3 2 Dimensional Code Reader Fixed Type Operation Manual Produced August 2001...
Страница 7: ......
Страница 9: ......
Страница 17: ......
Страница 27: ......
Страница 47: ......
Страница 63: ......
Страница 224: ...207 System Section 6 4 2 Set each item 3 Select End before leaving this screen...
Страница 226: ...209 System Section 6 4 2 Select the error method 3 Select End to leave this screen...
Страница 247: ...230 2 Dimensional Code Reader Section 8 1 8 1 2 Dimensional Code Reader V530 R150 Four 4 5 dia 165 min Unit mm...
Страница 263: ......
Страница 279: ......
Страница 281: ......
Страница 283: ......
Страница 289: ......