21
Figure 5-9. Analog Output Option Board
5.2.4 PANEL-MOUNT ASSEMBLY
The meter can be mounted in a panel so that the front of the meter is flush with the vertical panel
surface. Panel mounting can be seen as simply “sandwiching” the panel between the inner case and
the outer sleeve in the last phases of assembly.
Figure 5-10
shows the panel cutout dimensions, and
the dimensions for the panel thickness. It requires that the following already be done:
1.
Your line voltage rating has been checked against the meter rating on the Product ID and
serial number label on the meter case. See
Section 5.2.1
.
2.
You have configured all jumpers–those on the main board as well as those on any optional
boards. See
Section 5.2.2
for main board jumper configuration and the appropriate sections
for optional board jumper configuration.
3.
You have installed all optional boards and inserted the main board assembly back into the
case. See
Section 5.2.3
.
4.
You have wired P1, the AC power connector, and P2 the input output control connector;
connectors are not installed in the meter, but are ready to be installed. See
Section 7
.
5.
You have wired all connectors for optional boards; connectors are not connected to the meter,
but are ready to be installed.
If all of these steps are done, use
Figure 5-10
as a guide:
CAUTION:
Connectors with the wiring connections will be installed after mounting the unit.
Setup
5
S1
A
1
0
J5
1
1
1A
S1
P12
S1-A must be closed.
Refer to your Analog Output Option
Manual for more details.